Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: ECD 5100: CAL ADJ /5500 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 2 NUMBER OF LINES: 105 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- J F W 1.003 ASK+ K 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Connect UUT to AC power and turn unit ON. 1.006 DISP Warm-up time: 15 minutes 1.006 DISP Temperature: 0 -50degC 1.006 DISP Humidity: 30% - 60% 1.007 DISP [32] WARNING 1.007 DISP HIGH VOLTAGE is used or exposed during the performance 1.007 DISP of this calibration. DEATH ON CONTACT may result if 1.007 DISP personnel fail to observe safety precautions. 1.008 DISP This procedure was written to test the 5000-series 1.008 DISP dataloggers with a TV-10 or TV-40 MUX CARD installed. 1.009 HEAD MAX CHANNEL CONFIGURATION 1.010 DISP NOTE When programming data into the 5000-series 1.010 DISP datalogger, be certain that a "beep" is heard when 1.010 DISP pressing a key to insure the CPU has responded to 1.010 DISP the key stroke. 1.011 DISP Press the key to select alternate. 1.011 DISP Press the <9> key to select the MAX channels. 1.011 DISP Press the <0>, then <2> to select two channels. 1.012 HEAD DWELL CONFIGURATION 1.013 DISP Press the key to select alternate. 1.013 DISP Press the <6> key to select the dwell time. 1.013 DISP Press the <0>, then <5> to set dwell to 5sec. 1.014 HEAD RESET M & B PROGRAMMING CONSTANTS 1.015 DISP Press the key to select alternate. 1.015 DISP Press the key to select M coefficient. 1.015 DISP Press the <+>,<1>,<0>,<0>,<0>,<0> keys to set the 1.015 DISP M constant to +1.0000. 1.016 DISP Press the key to select alternate. 1.016 DISP Press the key to select B coefficient. 1.016 DISP Press the <+>,<0>,<0>,<0>,<0>,<0> keys to set the 1.016 DISP the B constant to +0.0000. 1.017 HEAD SET CHANNEL 1 TO READ VOLTAGE 1.018 DISP Press the key to select alternate. 1.018 DISP Press the key to select voltage. 1.018 DISP Press the <0>, then <1> to set the channel to 1. 1.019 HEAD SET CHANNEL 2 TO READ THERMOCOUPLE (J-type) 1.020 DISP Press the key to select alternate. 1.020 DISP Press the key to select J-type thermocouple. 1.020 DISP Press the <0>, then <2> to set the channel to 2. 1.021 HEAD SET CHANNEL 2 TADJUST 1.022 DISP Press the key to select alternate. 1.022 DISP Press the key to select Tadjust. 1.022 DISP Press the <+>,<0>,<0>,<0> keys to set the 1.022 DISP Tadjust to +0000.0. 1.023 HEAD DC VOLTAGE 1.024 DISP Connect the 5500A to the TV-10(TV-40)MUX as follows: 1.024 DISP [32] 5500A NORMAL HI to CHANNEL 1 HI 1.024 DISP [32] 5500A NORMAL LO to CHANNEL 1 LO 1.025 DISP HOLD the datalogger on channel 1 by pressing the 1.025 DISP ,,<0>,<1> keys. 1.026 5500 2.0000V S 2W 1.027 DISP Adjust R32 (top left of MUX card) for a reading 1.027 DISP of EXACTLY 2.0000. 1.028 5500 * S 1.029 HEAD TEMPERATURE 1.030 DISP Connect the 5500A to the TV-10(TV-40)MUX as follows: 1.030 DISP [32] 5500A NORMAL HI to CHANNEL 2 HI 1.030 DISP [32] 5500A NORMAL LO to CHANNEL 2 LO 1.031 DISP HOLD the datalogger on channel 2 by pressing 1.031 DISP the ,,<0>,<2> keys. 1.032 DISP Set SW1-1 (on MUX card) to the "ON" position. This 1.032 DISP removes the compensation from temperature readings. 1.033 5500 35.999mV S 2W 1.034 DISP Adjust R9 (top right of MUX card) for a reading 1.034 DISP of EXACTLY 1200.0 degrees F. 1.035 5500 * S 1.036 DISP Set SW1-1 (on MUX card) to the "OFF" position. 1.037 HEAD COLD JUNCTION COMPENSATION CHECK 1.038 DISP Short the CHANNEL 2 INPUT using a copper wire. 1.039 EVAL Does UUT read AMBIENT TEMPERATURE +/-5degF. 2.001 END